Calibration of lock-in amplifiers in the low-frequency range
calibration,linearity,demodulation,measurement techniques,voltage measurement,signal detection,Lock-in amplifier,gain error
摘要录用
Alessandro Cultrera / INRIM - Istituto Nazionale di Ricerca Metrologica
Ngoc Thanh Mai Tran / Politecnico di Torino
Vincenzo D'Elia / INRIM - Istituto Nazionale di Ricerca Metrologica
Massimo Ortolano / Politecnico di Torino
Luca Callegaro / INRIM - Istituto Nazionale di Ricerca Metrologica
A setup for the calibration of lock-in amplifiers in the low frequency range is presented. The setup is based on an uncalibrated voltage source, an ac voltmeter, and calibrated inductive and resistive voltage dividers. The measurand is the relative deviation of the lock-in amplifier reading from the calibrated applied voltage. Examples of calibration of a commercial lock-in amplifier in the 10 µV range, at frequencies of 17 Hz and 103 Hz, are given; the calibration uncertainty, relative to the instrument full-scale, is under evaluation and expected to be of the order of 10-4.